Induction

From Bangor Imaging Unit

Level 2 operators can provide the Induction for level 1 certification and must, at a minimum, cover the following:

Level 1 Induction

  • Screen the individual for MR contraindications and enter their details electronically into database
  • Have them read the BIU Standard Operating Procedure, and the MRI Operating Rules (available on the WIKI)
  • Have them view the 3 Phillips safety videos on the Wiki
  • One to one review of three key hazards:
    • Magnetic objects and other objects (e.g., pacemakers) vulnerable to magnets (highlight zone 1 and zone 2 and keeping doors closed, etc)
    • Quenching (what it is, when and associated hazards, inc. oxygen depletion and burning, noting oxygen monitor)
    • Noise-related safety
  • Discuss coiling of wires
  • Discuss hand clasping and leg crossing
  • Highlight evacuation procedure, fire exits, and location of fire extinguishers (MR safe versus not).
  • Review role of person B in emergency and fire-related emergencies
  • Demonstrate getting subject in and out of scanner
  • They also need to have the trainee complete the safety quiz (available from the level 1 safety officer) and see that this is done without access to notes.